Line 4: Line 4:


* "TV Mini-SEM (R)" by ISI (International Scientific Instruments)
* "TV Mini-SEM (R)" by ISI (International Scientific Instruments)
* lives in the room that used to house the laser cutter
* lives in the "science ghetto" on the west end of the south wall at 2169.
* is tended by Alan Rockefeller, Jeff Miller, Chris Murphy and Jim and the rest of the [https://www.noisebridge.net/mailman/listinfo/sem SEM mailing list]
* is tended by Mike Kan and the rest of the [https://www.noisebridge.net/mailman/listinfo/sem SEM mailing list]


=== Warnings ===
=== Warnings ===
# If you release the vacuum without venting the pump hose, pump oil will flood the chamber and contaminate the entire mechanism, requiring 10 hours of cleanup.
# If you release the vacuum without venting the pump hose, pump oil will flood the chamber and contaminate the entire mechanism, requiring 10 hours of cleanup.
# If you vent air into the column without powering down the filament, you'll probably destroy the filament ($100). The filament has a design lifetime of about 25 hours, depending on the current put through it. Adjust the specimen stage and other controls at low current, reserve higher currents for final photos.
# If you vent air into the column without powering down the filament, you'll probably destroy the filament ($100).
# Mike Kan has warned that if anyone asks him anything about the SEM, he's going to move it to another hackerspace. Jeff Miller has a MinI-SEM of his own, and knows its ins and outs pretty well.
# If you


=== Startup procedures ===
=== Startup procedures ===
# If the power cord is not already present, ask around or post to the list a few days prior to using the 'scope.
# get the power cord from Mike.
# See PDFs on vacuum start up and operators manual linked below.
# (to be continued)


=== Todo list ===
=== Todo list ===
Short Term:
# column liner needs to be replaced. Larry may be able to provide a clean liner. (2010-06-15)
# Isolate vibration from external sources; buses, people, etc. Mostly finished. (2013-6-20)
# (long term) build digital capture stage to replace CRT mechanism (2010-06-15)
# Figure out what to do about the person(s) now "occupying" the clean room which houses the SEM. (2013-6-20)
# build sputter coater for specimen preparation (2010-06-15)
# Design/build/find a sputter coater so the SEM can be used as a useful tool as well as a curiosity.
# test repair of X Stigmator circuit. Replaced 741 op amp. (ongoing, work began 2010-11-20)
# Create/obtain specimen stage clamp or spring to prevent specimen vibration. (2010-12-12)
# Obtain some sputter coated specimens. (2010-12-12)


Medium Term:
=== Repair and Enhancement ===
# Build a forepump line vibration isolator. (2013-6-20)
# Relocate or otherwise isolate the forepump. (2013-6-20)
# Replace most of the rubber tubing with metal tubing to reduce outgassing and pump-down times. (2013-6-20)
 
Long Term:
# Column liner needs to be replaced.  Larry may be able to provide a clean liner. (2010-06-15) (update: Larry has taken column liner to his lab for cleaning)(update: column liner is now clean) (2013-6-20)
# Integrate the digital capture and control unit Jeff built for his own Mini-SEM into the unit. (2013-6-20)
# Test repair of X Stigmator circuit. Replaced 741 op amp. (completed, work began 2010-11-20, tested 2010-12-10)
# Adjust vertical overscan of CRT (completed 2010-12-12, including hack sawing access port to overscan potentiometers in the CRT module)
# Fix 2x magnification toggle switch (2010-12-12)
# Diagnose and repair barrel distortion and occlusion at low magnification (2010-12-12)
# Sew dust cover (2010-12-12). Sew/heat seam new plastic dust cover, evidently the previous dust cover was destroyed. (2013-6-20)
# Rubber grommet around edge of CRT potentiometer port needed.
# Replace forepump with a (hopefully larger) pump that has a ballast port.
# Buy/find some real vacuum valves, turn simple spool pieces to serve as adapters, and electro-pneumatically interlock the valves so that valve sequencing is fool-proof.
# Evaluate air-cooled diffusion pump as source of vibration and contamination. Consider Santovac fluid and/or water-cooled pump and/or thermoelectric cold cap.
# Consider replacing filament with LaB6 cathode for brighter, sharper images and lower long-term per hour operating costs.
 
=== Repair and Enhancement Notes===
* Before repair, disable pumps and high voltage then put it into operation mode by placing a magnet on the operation relay.
* schematics framed behind SEM are to a similar, potentially slightly more modern model
* schematics framed behind SEM are to a similar, potentially slightly more modern model
* poorly designed wire harness may cause linear op amp and transistor on the J1 board of the power module to fail.    
* poorly designed wire harness may cause linear op amp and transistor on the J1 board of the CRT module to fail   
* camera mounting stage and bellows hood for slow-scan exposure. Proposed (2010-06-15), Completed by Mike, Miloh, and Chris (2010-11-06)
* camera mounting stage and bellows hood for slow-scan exposure. Proposed (2010-06-15), Completed by Mike, Miloh, and Chris (2010-11-06)
* CRT V. scan, V bias, H. frequency potentiometers are located on the underside of the CRT module. 
* broken stigmator Y lock tab is in altoids tin in SEM kit box.
How to mitigate vibration:
# Measure and pinpoint vibration sources
# Dampen vibration from the rough pump-line by one of these methods: routing it through a concrete block and adding a hose extension,  sandbag the existing line, or utilize water as a mass.
# Hang the column from the ceiling with bungee cord and platform
# Potentially relocate the microscope to the server room to diminish dust and increase distance from Mission St.
# Create spring clamping specimen slug to keep the specimen from moving.


=== Specifications ===
=== Specifications ===
* Made in Japan
* Delivered in September, 1975
* Delivered in September, 1975
* magnifies up to 20,000x
* magnifies up to 20,000x (reliably up to 700x in its present condition)
* outputs to a CRT
* outputs to a CRT
* non-digital, non-NTSC, non-PAL, analog output signal
* non-digital, non-NTSC, non-PAL, analog output signal
Line 76: Line 44:
[[Media: Other_Docs.pdf]] - Repair notes
[[Media: Other_Docs.pdf]] - Repair notes


[[Media: Packing_Lists.pdf]] - Original Packing Lists, the contents of the shipping crate
Original Packing List - coming soon


=== Schematics ===
=== Schematics ===
 
[[Media: Polaron_Pirani_Gauge_Head_for_Sputter_Coater.pdf]]
Overall schematics for the TV MINI-SEM (Model: MRS-2-2 WN76-7-01). These schematics are for a model of SEM that is perhaps newer than than the SEM at Noisebridge. It is still an invaluable reference to the technician. Note: This schematic is a mosaic of 9 images stitched together with [http://hugin.sourceforge.net/ Hugin 2010.2.0]. Some minor alignment issues exist in the content of the schematic. I think alignment can be guessed/interpreted accurately.  Available in:
[[Media: TV_MINI-SEM_Schematics-Overall_MRS-2-2_WN76-7-01_Mosaic_v._1.0_Low-Res.jpg | LOW RES 1.5MB (4330 x 3109, posterized grayscale, JPEG)]] and
[[Media: TV_MINI-SEM_Schematics-Overall_MRS-2-2_WN76-7-01_Mosaic_v._1.0.jpg | HIGH RES 8MB (8680 x 6217, color, JPEG)]].


[[Media: Schematic_of_Photomultiplier.pdf‎]]
[[Media: Schematic_of_Photomultiplier.pdf‎]]


[[Media: Polaron_Pirani_Schematics-PE_Pir-SW_Issue_4_E5000.pdf]] - Polaron Pirani sputter coater schematics
Full schematics of the SEM - coming soon
 
[[Media: Polaron_Pirani_Gauge_Head_for_Sputter_Coater.pdf]]


=== Specimen Images ===
=== Specimen Images ===
 
[[Image:IMG_0004.jpg|thumb|left|Image of mechanical pencil graphite on an aluminum slug at approximately 20x magnification. The specimen is about 1 mm across. Higher magnification images of this specimen are not available. Imaged by Chris Murphy and Mike Kan.
<gallery widths=200px heights=166px>
]]
Image:IMG_0004.jpg|Image of mechanical pencil graphite smudge on an aluminum slug at approximately 20x magnification. The specimen is about 1 mm across. This is prior to repairing the X Stigmator. Magnification was limited. Imaged by Chris Murphy and Mike Kan.
 
Image:IMG_0652.jpg|Mechanical pencil graphite smudge on aluminum slug. 1000x magnification. This is post repair of X Stigmator. Imaged by the Noisebridge SEM. Mike Kan, Chris Murphy.
 
Image:IMG_0655.jpg| Tip of mechanical pencil graphite  on aluminum slug. 500x magnification. This is post repair of X Stigmator. Imaged by the Noisebridge SEM. Mike Kan, Chris Murphy.
 
Image:IMG_0654.jpg| Tip of mechanical pencil graphite on aluminum slug. 1000x magnification. This is post repair of X Stigmator. Imaged by the Noisebridge SEM. Mike Kan, Chris Murphy.
 
Image:IMG_0656.jpg| Tip of mechanical pencil graphite on aluminum slug. 5000x magnification. This is post repair of X Stigmator. Imaged by the Noisebridge SEM. Mike Kan, Chris Murphy.
 
Image:IMG_0657.jpg| Tip of mechanical pencil graphite on aluminum slug. 10,000x magnification. This is post repair of X Stigmator. Specimen was moving on the slug due to mechanical vibration. Imaged by the Noisebridge SEM. Mike Kan, Chris Murphy.
 
Image:IMG_0624.jpg|Semiconductor ball grid array at 20x magnification prior to repairing the X Stigmator. Mike Kan, Chris Murphy
 
Image:IMG_0649.jpg|Snapped metal region of a box cutter razor blade. Imaged on the Noisebridge SEM. About 5000x magnification. This is post repair of X Stigmator. Mike Kan, Chris Murphy
 
Image:IMG_0007.jpg|Head of a small 2mm allen type machine screw. Less than 1000x magnification. This is post repair of X Stigmator. Mike Kan, Chris Murphy
 
Image:Allen_close_up.jpg|Head of a small 2mm allen type machine screw. About 1000x magnification. This is post repair of X Stigmator. Mike Kan, Chris Murphy
 
Image:IMG_0026.jpg|Diatom imaged in fast scan mode (lower resolution than slow scan photo mode). Magnification unrecorded. Mike Kan, Chris Murphy. 12/12/10
 
</gallery>
 
[[Category:Pages with a Noisebridge Tiny URL]]
Please note that all contributions to Noisebridge are considered to be released under the Creative Commons Attribution-NonCommercial-ShareAlike (see Noisebridge:Copyrights for details). If you do not want your writing to be edited mercilessly and redistributed at will, then do not submit it here.
You are also promising us that you wrote this yourself, or copied it from a public domain or similar free resource. Do not submit copyrighted work without permission!

To protect the wiki against automated edit spam, we kindly ask you to solve the following CAPTCHA:

Cancel Editing help (opens in new window)